Home>ISO Standards>BS ISO 5838-1:2013 pdf free

BS ISO 5838-1:2013 pdf free

BS ISO 5838-1:2013 pdf free.Implants for surgery一Metallic skeletal pins and wires
This part of ISO 5838 specifies general requirements for metallic skeletal pins and wires for use in bone surgery, excluding wires for binding and twisting.
The following documents, in whole or in part, are normatively referenced in this document and are indispensable for its application. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 3651-2,Determination of resistance to intergranular corrosion of stainless steels一Part 2:Ferriti,austenitic and ferritic-austenitic (duplex) stainless steels 一Corrosion test in media containing sulfuric acid
ISO 5832-1, Implants for surgery一Metallic materials一Part 1: Wrought stainless steel
ISO 5832-2, Implants for surgery一Metallic materials一Part 2: Unalloyed titanium
ISO 5832-3, Implants for surgery一Metallic materials 一Part 3: Wrought titanium 6-aluminium 4-vanadium alloy
ISO 5832-5, Implants for surgery一Metallic materials一Part 5: Wrought cobalt-chromium-tungsten-nickel alloy
ISO 5832-6, Implants for surgery一Metallic materials- Part 6: Wrought cobalt-nickel-chromium-molybdenum alloy
ISO 5832-7, Implants for surgery一Metallic materials一Part 7: Forgeable and cold-formed cobalt-chromium-nickel-molybdenum-iron alloy
ISO 5832-8, Implants for surgery一Metallic materials- Part 8: Wrought cobalt-nickel-chromium-molybdenum- tungsten-iron alloy
ISO 5832-9, Implants for surgery一Metallic materials一Part 9: Wrought high nitrogen stainless steel
ISO 5832-11, Implants for surgery一Metallic materials 一Part 11: Wrought titanium 6-aluminium 7-niobium alloy
ISO 5832-12, Implants for surgery 一Metallic materials 一Part 12: Wrought cobalt-chromium-molybdenum alloy
ISO 5832-14, Implants for surgery一Metallic materials- Part 14: Wrought titanium 15- molybdenum 5-zirconium 3-aluminium alloy
ISO 6892-1, Metallic materials一Tensile testing一Part 1: Method of test at room temperature
ISO 14630:2012, Non-active surgical implants一General requirements
Skeletal pins and wires shall be manufactured by a cold drawing or a cold working process. Welds are not permissible. If local heating is needed for re-shaping points made of stainless steel, this can only be permitted provided that the whole of the final product can be shown to satisfy the intergranular corrosion test according to ISO 3651-2.
The external surface finish shall be free from burs, scratches and other defects visible to the naked eye.The products shall be appropriately treated to provide surface passivation. Special attention should be paid to screw threads, if applicable.BS ISO 5838-1 pdf free.

Related standards